| | EMC Directive, RoHS Directive | Sealed gauge pressure (Note): The sensor's internal mechanism is sealed based on an air pressure of 1,013 hPa. | 0 to +3.5 MPa | | | 7 MPa | Gases and fluids that do not corrode stainless steel SUS630, SUS304, or SUSXM7 | | | 9 to 36 V DC [9 to 32 V DC when using the attached connector (e-CON)] | | 20 mA or less | Output voltage: 1 to 5 V DC (over rated pressure range)
Accuracy: ±1.0 % F.S. (at 23 ±2 ℃ +73.4 ±35 ℉)
±2.0 % F.S. (at -20 to 70 ℃ -4 to +158 ℉)
(including linearity, hysteresis and repeatability) | 1 ms or less | | | | | | | | | | | | IP67 (IEC) | -20 to +70 ℃ -4 to +158 ℉ (No dew condensation or icing allowed),
Storage: -30 to +70℃ -22 to +158 ℉ | 35 to 85 % RH, Storage : 35 to 85 % RH | -20 to +70 ℃ -4 to +158 ℉ | 150 V AC for one min. between all supply terminals connected together and enclosure | 100 MΩ, or more, with 50 V DC megger between all supply terminals connected together and enclosure | 10 to 2,000 Hz frequency, acceleration 200 m/s 2, in X direction for four hours, in Y and Z directions for two hours each (Note): The X, Y, and Z directions are defined as follows :  | 1,000 m/s2 acceleration in X, Y and Z directions three times each | | Capacitor earth (Enclosure-supply terminal) | R1/4 male thread (throttle embeded) | Diaphragm : Stainless steel (SUS603), Pressure port : Stainless steel (SUS304), Throttle : Stainless steel (SUSXM7) | Connector | | 0.2 mm2 3-core heat resistant cabtyre cable 2 m 3.562 ft long | Extension up to total 10 m 32.808 ft is possible with 0.2 mm2, or more, cable. | Net weight : 100 g approx., Gross weight : 150 g approx. | Connector (e-CON) : 1 pc. | Where measurement conditions have not been specified precisely, the conditions used were an ambient temperature of +23 ℃ +73.4 ℉. The sensor head can be used independently. | CE Marking (EMC Directive, RoHS Directive), UKCA Marking (EMC Regulations, RoHS Regulations) |
| | EMC Directive, RoHS Directive | Sealed gauge pressure (Note): The sensor's internal mechanism is sealed based on an air pressure of 1,013 hPa. | 0 to +50 MPa | | | 75 MPa | Gases and fluids that do not corrode stainless steel SUS630, SUS304, or SUSXM7 | | | 9 to 36 V DC [9 to 32 V DC when using the attached connector (e-CON)] | | 20 mA or less | Output voltage: 1 to 5 V DC (over rated pressure range)
Accuracy: ±1.0 % F.S. (at 23 ±2 ℃ +73.4 ±35 ℉)
±2.0 % F.S. (at -20 to 125 ℃ -4 to +257 ℉)
(including linearity, hysteresis and repeatability) | 1 ms or less | | | | | | | | | | | | IP67 (IEC) | -20 to +80 ℃ (Pressure port: -20 to +125 ℃
-4 to +257 ℉, No dew condensation or icing allowed),
Storage: -30 to +100 ℃ -22 to +212 ℉
[e-CON connector (accesory): -20 to +75 ℃
-4 to +167 ℉
(Storage: -30 to +75 ℃ -22 to +167 ℉)] | 35 to 85 % RH, Storage : 35 to 85 % RH | -20 to +125 ℃ -4 to +257 ℉ | 150 V AC for one min. between all supply terminals connected together and enclosure | 100 MΩ, or more, with 50 V DC megger between all supply terminals connected together and enclosure | 10 to 2,000 Hz frequency, acceleration 200 m/s 2, in X direction for four hours, in Y and Z directions for two hours each (Note): The X, Y, and Z directions are defined as follows :  | 1,000 m/s2 acceleration in X, Y and Z directions three times each | | Capacitor earth (Enclosure-supply terminal) | R1/4 male thread (throttle embeded) | Diaphragm : Stainless steel (SUS603), Pressure port : Stainless steel (SUS304), Throttle : Stainless steel (SUSXM7) | Connector | | 0.2 mm2 3-core heat resistant cabtyre cable 2 m 3.562 ft long | Extension up to total 10 m 32.808 ft is possible with 0.2 mm2, or more, cable. | Net weight : 100 g approx., Gross weight : 150 g approx. | Connector (e-CON) : 1 pc. | Where measurement conditions have not been specified precisely, the conditions used were an ambient temperature of +23 ℃ +73.4 ℉. The sensor head can be used independently. Oil is used in the factory inspection process for models DPH-L114 and DPH-L154. There may be some residual oil inside the pressure port. | CE Marking (EMC Directive, RoHS Directive), UKCA Marking (EMC Regulations, RoHS Regulations) |
| | EMC Directive, RoHS Directive | Sealed gauge pressure (Note): The sensor's internal mechanism is sealed based on an air pressure of 1,013 hPa. | 0 to +10 MPa | | | 20 MPa | Gases and fluids that do not corrode stainless steel SUS630, SUS304, or SUSXM7 | | | 9 to 36 V DC [9 to 32 V DC when using the attached connector (e-CON)] | | 20 mA or less | Output voltage: 1 to 5 V DC (over rated pressure range)
Accuracy: ±1.0 % F.S. (at 23 ±2 ℃ +73.4 ±35 ℉)
±2.0 % F.S. (at -20 to 125 ℃ -4 to +257 ℉)
(including linearity, hysteresis and repeatability) | 1 ms or less | | | | | | | | | | | | IP67 (IEC) | -20 to +80 ℃ (Pressure port: -20 to +125 ℃
-4 to +257 ℉, No dew condensation or icing allowed),
Storage: -30 to +100 ℃ -22 to +212 ℉
[e-CON connector (accesory): -20 to +75 ℃
-4 to +167 ℉
(Storage: -30 to +75 ℃ -22 to +167 ℉)] | 35 to 85 % RH, Storage : 35 to 85 % RH | -20 to +125 ℃ -4 to +257 ℉ | 150 V AC for one min. between all supply terminals connected together and enclosure | 100 MΩ, or more, with 50 V DC megger between all supply terminals connected together and enclosure | 10 to 2,000 Hz frequency, acceleration 200 m/s 2, in X direction for four hours, in Y and Z directions for two hours each (Note): The X, Y, and Z directions are defined as follows :  | 1,000 m/s2 acceleration in X, Y and Z directions three times each | | Capacitor earth (Enclosure-supply terminal) | R1/4 male thread (throttle embeded) | Diaphragm : Stainless steel (SUS603), Pressure port : Stainless steel (SUS304), Throttle : Stainless steel (SUSXM7) | Connector | | 0.2 mm2 3-core heat resistant cabtyre cable 2 m 3.562 ft long | Extension up to total 10 m 32.808 ft is possible with 0.2 mm2, or more, cable. | Net weight : 100 g approx., Gross weight : 150 g approx. | Connector (e-CON) : 1 pc. | Where measurement conditions have not been specified precisely, the conditions used were an ambient temperature of +23 ℃ +73.4 ℉. The sensor head can be used independently. Oil is used in the factory inspection process for models DPH-L114 and DPH-L154. There may be some residual oil inside the pressure port. | CE Marking (EMC Directive, RoHS Directive), UKCA Marking (EMC Regulations, RoHS Regulations) |
| | EMC Directive, RoHS Directive | Sealed gauge pressure (Note): The sensor's internal mechanism is sealed based on an air pressure of 1,013 hPa. | 0 to +1 MPa | | | 2 MPa | Gases and fluids that do not corrode stainless steel SUS630, SUS304, or SUSXM7 | | | 9 to 36 V DC [9 to 32 V DC when using the attached connector (e-CON)] | | 20 mA or less | Output voltage: 1 to 5 V DC (over rated pressure range)
Accuracy: ±1.0 % F.S. (at 23 ±2 ℃ +73.4 ±35 ℉)
±2.0 % F.S. (at -20 to 70 ℃ -4 to +158 ℉)
(including linearity, hysteresis and repeatability) | 1 ms or less | | | | | | | | | | | | IP67 (IEC) | -20 to +70 ℃ -4 to +158 ℉ (No dew condensation or icing allowed),
Storage: -30 to +70℃ -22 to +158 ℉ | 35 to 85 % RH, Storage : 35 to 85 % RH | -20 to +70 ℃ -4 to +158 ℉ | 150 V AC for one min. between all supply terminals connected together and enclosure | 100 MΩ, or more, with 50 V DC megger between all supply terminals connected together and enclosure | 10 to 2,000 Hz frequency, acceleration 200 m/s 2, in X direction for four hours, in Y and Z directions for two hours each (Note): The X, Y, and Z directions are defined as follows :  | 1,000 m/s2 acceleration in X, Y and Z directions three times each | | Capacitor earth (Enclosure-supply terminal) | R1/4 male thread (throttle embeded) | Diaphragm : Stainless steel (SUS603), Pressure port : Stainless steel (SUS304), Throttle : Stainless steel (SUSXM7) | Connector | | 0.2 mm2 3-core heat resistant cabtyre cable 2 m 3.562 ft long | Extension up to total 10 m 32.808 ft is possible with 0.2 mm2, or more, cable. | Net weight : 100 g approx., Gross weight : 150 g approx. | Connector (e-CON) : 1 pc. | Where measurement conditions have not been specified precisely, the conditions used were an ambient temperature of +23 ℃ +73.4 ℉. The sensor head can be used independently. | CE Marking (EMC Directive, RoHS Directive), UKCA Marking (EMC Regulations, RoHS Regulations) |
| | | | DPH-L113V : -0.1 to +1 MPa DPH-L113 : 0 to +1 MPa DPH-L133 : 0 to +3.5 MPa DPH-L114 : 0 to +10 MPa DPH-L154 : 0 to +50 MPa (Note) : Values differ depending on a connected sensor head. | [DPH-L113V]
-1.177 to +1.177 MPa
-12.00 to +12.00 kgf/cm2
-11.77 to +11.77 bar
-170.6 to +170.6 psi
[DPH-L113]
-1.070 to +1.070 MPa
-10.91 to +10.91 kgf/cm2
-10.70 to +10.70 bar
-155.2 to +155.2 psi
[DPH-L133]
-3.74 to +3.74 MPa
-38.1 to +38.1 kgf/cm2
-37.4 to +37.4 bar
-542 to +542 psi
[DPH-L114]
-10.70 to +10.70 MPa
-109.1 to +109.1 kgf/cm2
-107.0 to +107.0 bar
-1552 to +1552 psi
[DPH-L154]
-53.5 to +53.5 MPa
-545 to +545 kgf/cm2
-535 to +535 bar
-1980 to +7760 psi
(Note) : Values differ depending on a connected sensor head. | DPH-L113V : 0.001 MPa DPH-L113 : 0.001 MPa DPH-L133 : 0.01 MPa DPH-L114 : 0.01 MPa DPH-L154 : 0.1 MPa (Note): Values differ depending on a connected sensor head. | | | 4 digits +4 digits 3-color LCD display (Display refresh rate : 250 ms, 500 ms, 1,000 ms, selectable by key operation) | [DPH-L113V]
-0.155 to +1.022 MPa
-1.58 to +10.42 kgf/cm2
-1.55 to +10.22 bar
-22.4 to +148.2 psi
[DPH-L113]
-0.050 to +1.020 MPa
-0.51 to +10.40 kgf/cm2
-0.50 to +10.20 bar
-7.2 to +148.0 psi
[DPH-L133]
-0.17 to +3.57 MPa
-1.7 to +36.4 kgf/cm2
-1.7 to +35.7 bar
-24 to +518 psi
[DPH-L114]
-0.50 to +10.20 MPa
-5.1 to +104.0 kgf/cm2
-5.0 to +102.0 bar
-72 to +1480 psi
[DPH-L154]
-2.5 to +51.0 MPa
-25 to +520 kgf/cm2
-25 to +510 bar
-360 to +7400 psi
(Note): Values differ depending on a connected sensor head. | 12 to 24 V DC ±10 % Ripple P-P 10 % or less | Normal operation : 960 mW or less (Current consumption 40 mA or less at 24 V supply voltage) ECO mode (STD) : 720 mW or less (Current consumption 30 mA or less at 24 V supply voltage) ECO mode (FULL) : 600 mW or less (Current consumption 25 mA or less at 24 V supply voltage) Excluding the current consumption of sensor head and analog output current | | | | NPN open-collector transistor (2 outputs) ・ Maximum sink current : 50 mA ・ Applied voltage : 30 V DC or less (between comparative output and 0 V) ・ Residual voltage : 1 V or less (at 50 mA sink current)
| NO / NC, selectable by key operation | EASY mode / Hysteresis mode / Window comparator mode | Minimum 1 digit (variable)(however, 2 digits when using psi unit) | Within ±0.2 % F.S. | 5 ms, 10 ms, 25 ms, 50 ms, 100 ms, 250 ms, 500 ms, 1,000 ms, 5,000 ms, selectable by key operation | Incorporated | ・ Output current : 1 to 5 V DC ・ Zero point : within 1 V ±0.5 % F.S. (excluding DPH-L113V) within 1.364 V ±0.5 % F.S. (DPH-L113V) ・ Span : within 4 V ±0.5 % F.S. ・ Linearity : within ±0.1 % F.S. ・ Output impedance : 1 kΩ approx.
・ Output current : 4 to 20 mA ・ Zero point : within 4 mA ±1.0 % F.S. (excluding DPH-L113V) within 5.455 mA ±1.0 % F.S. (DPH-L113V) ・ Span : within 16 mA ±1.5 % F.S. ・ Linearity : within ±0.1 % F.S. ・ Load resistance : 250 Ω (max.)
| Input voltage range : 1 to 5 V DC (over rated pressure range) | ON voltage : 0.4 V DC or less OFF voltage : 5 to 30 V DC, or open Input impedance : 10 kΩ approx. Input time : 1 ms or more
| Orange LED (Comparative output 1 operation indicator, comparative output 2 operation indicator : Lights up when each comparative output is ON ) | IP40 (IEC) | -10 to +50 ℃ +14 to +122 ℉ (No dew condensation or icing allowed), Storage : -10 to +60 ℃ +14 to +140 ℉ | 35 to 85 % RH, Storage : 35 to 85 % RH | | 500 V AC for one min. between all supply terminals connected together and enclosure | 50 MΩ, or more, with 500 V DC megger between all supply terminals connected together and enclosure | 10 to 500 Hz frequency, amplitude 3 mm 0.118 in or maximum acceleration 196 m/s2, in X, Y and Z directions for two hours each (when panel mounting bracket is mounted : 10 to 150 Hz frequency, amplitude 0.75 mm 0.030 in or maximum acceleration 49 m/s2, in X, Y and Z directions for two hours each) | 100 m/s2 acceleration in X, Y and Z directions for three times each | Within ±0.5 % F.S. (ambient temperature range based on +20 ℃ +68 ℉) | | | Enclosure : PBT (glass fiber reinforced), LCD display : Acrylic, Mounting threaded part : Brass (nickel plated), Switch part : Silicone rubber | Connector | Total length up to 100 m 328.1 ft (less than 30 m 98.4 ft when conforming to CE marking) is possible with 0.3 mm2, or more, cable. | | | Net weight : 25 g approx. (excluding connector attached cable), Gross weight : 140 g approx. | Pressure unit label: 1 set | Where measurement conditions have not been specified precisely, the conditions used were an ambient temperature of +20 ℃ +68 ℉. The values specified above are applied only to the controller. | |
| | EMC Directive, RoHS Directive | | DPH-L113V : -0.1 to +1 MPa DPH-L113 : 0 to +1 MPa DPH-L133 : 0 to +3.5 MPa DPH-L114 : 0 to +10 MPa DPH-L154 : 0 to +50 MPa (Note) : Values differ depending on a connected sensor head. | [DPH-L113V] -1.177 to +1.177 MPa -12.00 to +12.00 kgf/cm2 -11.77 to +11.77 bar -170.6 to +170.6 psi [DPH-L113] -1.070 to +1.070 MPa -10.91 to +10.91 kgf/cm2 -10.70 to +10.70 bar -155.2 to +155.2 psi [DPH-L133] -3.74 to +3.74 MPa -38.1 to +38.1 kgf/cm2 -37.4 to +37.4 bar -542 to +542 psi [DPH-L114] -10.70 to +10.70 MPa -109.1 to +109.1 kgf/cm2 -107.0 to +107.0 bar -1552 to +1552 psi [DPH-L154] -53.5 to +53.5 MPa -545 to +545 kgf/cm2 -535 to +535 bar -1980 to +7760 psi (Note) : Values differ depending on a connected sensor head. | DPH-L113V : 0.001 MPa DPH-L113 : 0.001 MPa DPH-L133 : 0.01 MPa DPH-L114 : 0.01 MPa DPH-L154 : 0.1 MPa (Note): Values differ depending on a connected sensor head. | | | 4 digits +4 digits 3-color LCD display (Display refresh rate : 250 ms, 500 ms, 1,000 ms, selectable by key operation) | [DPH-L113V] -0.155 to +1.022 MPa -1.58 to +10.42 kgf/cm2 -1.55 to +10.22 bar -22.4 to +148.2 psi [DPH-L113] -0.050 to +1.020 MPa -0.51 to +10.40 kgf/cm2 -0.50 to +10.20 bar -7.2 to +148.0 psi [DPH-L133] -0.17 to +3.57 MPa -1.7 to +36.4 kgf/cm2 -1.7 to +35.7 bar -24 to +518 psi [DPH-L114] -0.50 to +10.20 MPa -5.1 to +104.0 kgf/cm2 -5.0 to +102.0 bar -72 to +1480 psi [DPH-L154] -2.5 to +51.0 MPa -25 to +520 kgf/cm2 -25 to +510 bar -360 to +7400 psi (Note): Values differ depending on a connected sensor head. | 12 to 24 V DC ±10 % Ripple P-P 10 % or less | Normal operation : 960 mW or less (Current consumption 40 mA or less at 24 V supply voltage) ECO mode (STD) : 720 mW or less (Current consumption 30 mA or less at 24 V supply voltage) ECO mode (FULL) : 600 mW or less (Current consumption 25 mA or less at 24 V supply voltage) Excluding the current consumption of sensor head and analog output current | | | | PNP open-collector transistor (2 outputs) ・ Maximum source current : 50 mA ・ Applied voltage : 30 V DC or less (between comparative output and +V) ・ Residual voltage : 1 V or less (at 50 mA source current)
| NO / NC, selectable by key operation | EASY mode / Hysteresis mode / Window comparator mode | Minimum 1 digit (variable)(however, 2 digits when using psi unit) | Within ±0.2 % F.S. | 5 ms, 10 ms, 25 ms, 50 ms, 100 ms, 250 ms, 500 ms, 1,000 ms, 5,000 ms, selectable by key operation | Incorporated | ・ Output current : 1 to 5 V DC ・ Zero point : within 1 V ±0.5 % F.S. (excluding DPH-L113V) within 1.364 V ±0.5 % F.S. (DPH-L113V) ・ Span : within 4 V ±0.5 % F.S. ・ Linearity : within ±0.1 % F.S. ・ Output impedance : 1 kΩ approx.
・ Output current : 4 to 20 mA ・ Zero point : within 4 mA ±1.0 % F.S. (excluding DPH-L113V) within 5.455 mA ±1.0 % F.S. (DPH-L113V) ・ Span : within 16 mA ±1.5 % F.S. ・ Linearity : within ±0.1 % F.S. ・ Load resistance : 250 Ω (max.)
| Input voltage range : 1 to 5 V DC (over rated pressure range) | ON voltage : 5 V to +V DC OFF voltage : 0.6 V DC or less, or open Input impedance : 10 kΩ approx. Input time : 1 ms or more
| Orange LED (Comparative output 1 operation indicator, comparative output 2 operation indicator : Lights up when each comparative output is ON ) | IP40 (IEC) | -10 to +50 ℃ +14 to +122 ℉ (No dew condensation or icing allowed), Storage : -10 to +60 ℃ +14 to +140 ℉ | 35 to 85 % RH, Storage : 35 to 85 % RH | | 500 V AC for one min. between all supply terminals connected together and enclosure | 50 MΩ, or more, with 500 V DC megger between all supply terminals connected together and enclosure | 10 to 500 Hz frequency, double amplitude 3 mm 0.118 in or maximum acceleration 196 m/s2, in X, Y and Z directions for two hours each (when panel mounting bracket is mounted: 10 to 150 Hz frequency, double amplitude 0.75 mm 0.030 in or maximum acceleration 49 m/s2, in X, Y and Z directions for two hours each) | 100 m/s2 acceleration in X, Y and Z directions three times each | Within ±0.5 % F.S. (ambient temperature range based on +20 ℃ +68 ℉) | | | Enclosure : PBT (glass fiber reinforced), LCD display : Acrylic, Mounting threaded part : Brass (nickel plated), Switch part : Silicone rubber | Connector | Total length up to 100 m 328.084 ft (less than 30 m 98.425 ft when conforming to CE marking) is possible with 0.3 mm2, or more, cable. | | | Net weight : 25 g approx. (excluding connector attached cable), Gross weight : 140 g approx. | CN-66A-C2 (Connector attached cable 2 m 6.562 ft), Pressure unit label: 1 set | Where measurement conditions have not been specified precisely, the conditions used were an ambient temperature of +20 ℃ +68 ℉. The values specified above are applied only to the controller. | CE Marking (EMC Directive, RoHS Directive), UKCA Marking (EMC Regulations, RoHS Regulations) |
| | EMC Directive, RoHS Directive | | DPH-L113V : -0.1 to +1 MPa DPH-L113 : 0 to +1 MPa DPH-L133 : 0 to +3.5 MPa DPH-L114 : 0 to +10 MPa DPH-L154 : 0 to +50 MPa (Note) : Values differ depending on a connected sensor head. | [DPH-L113V] -1.177 to +1.177 MPa -12.00 to +12.00 kgf/cm2 -11.77 to +11.77 bar -170.6 to +170.6 psi [DPH-L113] -1.070 to +1.070 MPa -10.91 to +10.91 kgf/cm2 -10.70 to +10.70 bar -155.2 to +155.2 psi [DPH-L133] -3.74 to +3.74 MPa -38.1 to +38.1 kgf/cm2 -37.4 to +37.4 bar -542 to +542 psi [DPH-L114] -10.70 to +10.70 MPa -109.1 to +109.1 kgf/cm2 -107.0 to +107.0 bar -1552 to +1552 psi [DPH-L154] -53.5 to +53.5 MPa -545 to +545 kgf/cm2 -535 to +535 bar -1980 to +7760 psi (Note) : Values differ depending on a connected sensor head. | DPH-L113V : 0.001 MPa DPH-L113 : 0.001 MPa DPH-L133 : 0.01 MPa DPH-L114 : 0.01 MPa DPH-L154 : 0.1 MPa (Note): Values differ depending on a connected sensor head. | | | 4 digits +4 digits 3-color LCD display (Display refresh rate : 250 ms, 500 ms, 1,000 ms, selectable by key operation) | [DPH-L113V] -0.155 to +1.022 MPa -1.58 to +10.42 kgf/cm2 -1.55 to +10.22 bar -22.4 to +148.2 psi [DPH-L113] -0.050 to +1.020 MPa -0.51 to +10.40 kgf/cm2 -0.50 to +10.20 bar -7.2 to +148.0 psi [DPH-L133] -0.17 to +3.57 MPa -1.7 to +36.4 kgf/cm2 -1.7 to +35.7 bar -24 to +518 psi [DPH-L114] -0.50 to +10.20 MPa -5.1 to +104.0 kgf/cm2 -5.0 to +102.0 bar -72 to +1480 psi [DPH-L154] -2.5 to +51.0 MPa -25 to +520 kgf/cm2 -25 to +510 bar -360 to +7400 psi (Note): Values differ depending on a connected sensor head. | 12 to 24 V DC ±10 % Ripple P-P 10 % or less | Normal operation : 960 mW or less (Current consumption 40 mA or less at 24 V supply voltage) ECO mode (STD) : 720 mW or less (Current consumption 30 mA or less at 24 V supply voltage) ECO mode (FULL) : 600 mW or less (Current consumption 25 mA or less at 24 V supply voltage) Excluding the current consumption of sensor head and analog output current | | | | NPN open-collector transistor (2 outputs) ・ Maximum sink current : 50 mA ・ Applied voltage : 30 V DC or less (between comparative output and 0 V) ・ Residual voltage : 1 V or less (at 50 mA sink current)
| NO / NC, selectable by key operation | EASY mode / Hysteresis mode / Window comparator mode | Minimum 1 digit (variable)(however, 2 digits when using psi unit) | Within ±0.2 % F.S. | 5 ms, 10 ms, 25 ms, 50 ms, 100 ms, 250 ms, 500 ms, 1,000 ms, 5,000 ms, selectable by key operation | Incorporated | ・ Output current : 1 to 5 V DC ・ Zero point : within 1 V ±0.5 % F.S. (excluding DPH-L113V) within 1.364 V ±0.5 % F.S. (DPH-L113V) ・ Span : within 4 V ±0.5 % F.S. ・ Linearity : within ±0.1 % F.S. ・ Output impedance : 1 kΩ approx.
・ Output current : 4 to 20 mA ・ Zero point : within 4 mA ±1.0 % F.S. (excluding DPH-L113V) within 5.455 mA ±1.0 % F.S. (DPH-L113V) ・ Span : within 16 mA ±1.5 % F.S. ・ Linearity : within ±0.1 % F.S. ・ Load resistance : 250 Ω (max.)
| Input voltage range : 1 to 5 V DC (over rated pressure range) | ON voltage : 0.4 V DC or less OFF voltage : 5 to 30 V DC, or open Input impedance : 10 kΩ approx. Input time : 1 ms or more
| Orange LED (Comparative output 1 operation indicator, comparative output 2 operation indicator : Lights up when each comparative output is ON ) | IP40 (IEC) | -10 to +50 ℃ +14 to +122 ℉ (No dew condensation or icing allowed), Storage : -10 to +60 ℃ +14 to +140 ℉ | 35 to 85 % RH, Storage : 35 to 85 % RH | | 500 V AC for one min. between all supply terminals connected together and enclosure | 50 MΩ, or more, with 500 V DC megger between all supply terminals connected together and enclosure | 10 to 500 Hz frequency, double amplitude 3 mm 0.118 in or maximum acceleration 196 m/s2, in X, Y and Z directions for two hours each (when panel mounting bracket is mounted: 10 to 150 Hz frequency, double amplitude 0.75 mm 0.030 in or maximum acceleration 49 m/s2, in X, Y and Z directions for two hours each) | 100 m/s2 acceleration in X, Y and Z directions three times each | Within ±0.5 % F.S. (ambient temperature range based on +20 ℃ +68 ℉) | | | Enclosure : PBT (glass fiber reinforced), LCD display : Acrylic, Mounting threaded part : Brass (nickel plated), Switch part : Silicone rubber | Connector | Total length up to 100 m 328.084 ft (less than 30 m 98.425 ft when conforming to CE marking) is possible with 0.3 mm2, or more, cable. | | | Net weight : 25 g approx. (excluding connector attached cable), Gross weight : 140 g approx. | CN-66A-C2 (Connector attached cable 2 m 6.562 ft), Pressure unit label: 1 set | Where measurement conditions have not been specified precisely, the conditions used were an ambient temperature of +20 ℃ +68 ℉. The values specified above are applied only to the controller. | CE Marking (EMC Directive, RoHS Directive), UKCA Marking (EMC Regulations, RoHS Regulations) |
| | EMC Directive, RoHS Directive | Sealed gauge pressure (Note): The sensor's internal mechanism is sealed based on an air pressure of 1,013 hPa. | -0.1 to +1 MPa | | | 2 MPa | Gases and fluids that do not corrode stainless steel SUS630, SUS304, or SUSXM7 | | | 9 to 36 V DC [9 to 32 V DC when using the attached connector (e-CON)] | | 20 mA or less | Output voltage: 1 to 5 V DC (over rated pressure range)
Accuracy: ±1.0 % F.S. (at 23 ±2 ℃ +73.4 ±35 ℉)
±2.0 % F.S. (at -20 to 70 ℃ -4 to +158 ℉)
(including linearity, hysteresis and repeatability) | 1 ms or less | | | | | | | | | | | | IP67 (IEC) | -20 to +70 ℃ -4 to +158 ℉ (No dew condensation or icing allowed),
Storage: -30 to +70℃ -22 to +158 ℉ | 35 to 85 % RH, Storage : 35 to 85 % RH | -20 to +70 ℃ -4 to +158 ℉ | 150 V AC for one min. between all supply terminals connected together and enclosure | 100 MΩ, or more, with 50 V DC megger between all supply terminals connected together and enclosure | 10 to 2,000 Hz frequency, acceleration 200 m/s 2, in X direction for four hours, in Y and Z directions for two hours each (Note): The X, Y, and Z directions are defined as follows :  | 1,000 m/s2 acceleration in X, Y and Z directions three times each | | Capacitor earth (Enclosure-supply terminal) | R1/4 male thread (throttle embeded) | Diaphragm : Stainless steel (SUS603), Pressure port : Stainless steel (SUS304), Throttle : Stainless steel (SUSXM7) | Connector | | 0.2 mm2 3-core heat resistant cabtyre cable 2 m 3.562 ft long | Extension up to total 10 m 32.808 ft is possible with 0.2 mm2, or more, cable. | Net weight : 100 g approx., Gross weight : 150 g approx. | Connector (e-CON) : 1 pc. | Where measurement conditions have not been specified precisely, the conditions used were an ambient temperature of +23 ℃ +73.4 ℉. The sensor head can be used independently. | CE Marking (EMC Directive, RoHS Directive), UKCA Marking (EMC Regulations, RoHS Regulations) |
| | | | DPH-L113V : -0.1 to +1 MPa DPH-L113 : 0 to +1 MPa DPH-L133 : 0 to +3.5 MPa DPH-L114 : 0 to +10 MPa DPH-L154 : 0 to +50 MPa (Note) : Values differ depending on a connected sensor head. | [DPH-L113V]
-1.177 to +1.177 MPa
-12.00 to +12.00 kgf/cm2
-11.77 to +11.77 bar
-170.6 to +170.6 psi
[DPH-L113]
-1.070 to +1.070 MPa
-10.91 to +10.91 kgf/cm2
-10.70 to +10.70 bar
-155.2 to +155.2 psi
[DPH-L133]
-3.74 to +3.74 MPa
-38.1 to +38.1 kgf/cm2
-37.4 to +37.4 bar
-542 to +542 psi
[DPH-L114]
-10.70 to +10.70 MPa
-109.1 to +109.1 kgf/cm2
-107.0 to +107.0 bar
-1552 to +1552 psi
[DPH-L154]
-53.5 to +53.5 MPa
-545 to +545 kgf/cm2
-535 to +535 bar
-1980 to +7760 psi
(Note) : Values differ depending on a connected sensor head. | DPH-L113V : 0.001 MPa DPH-L113 : 0.001 MPa DPH-L133 : 0.01 MPa DPH-L114 : 0.01 MPa DPH-L154 : 0.1 MPa (Note): Values differ depending on a connected sensor head. | | | 4 digits +4 digits 3-color LCD display (Display refresh rate : 250 ms, 500 ms, 1,000 ms, selectable by key operation) | [DPH-L113V]
-0.155 to +1.022 MPa
-1.58 to +10.42 kgf/cm2
-1.55 to +10.22 bar
-22.4 to +148.2 psi
[DPH-L113]
-0.050 to +1.020 MPa
-0.51 to +10.40 kgf/cm2
-0.50 to +10.20 bar
-7.2 to +148.0 psi
[DPH-L133]
-0.17 to +3.57 MPa
-1.7 to +36.4 kgf/cm2
-1.7 to +35.7 bar
-24 to +518 psi
[DPH-L114]
-0.50 to +10.20 MPa
-5.1 to +104.0 kgf/cm2
-5.0 to +102.0 bar
-72 to +1480 psi
[DPH-L154]
-2.5 to +51.0 MPa
-25 to +520 kgf/cm2
-25 to +510 bar
-360 to +7400 psi
(Note): Values differ depending on a connected sensor head. | 12 to 24 V DC ±10 % Ripple P-P 10 % or less | Normal operation : 960 mW or less (Current consumption 40 mA or less at 24 V supply voltage) ECO mode (STD) : 720 mW or less (Current consumption 30 mA or less at 24 V supply voltage) ECO mode (FULL) : 600 mW or less (Current consumption 25 mA or less at 24 V supply voltage) Excluding the current consumption of sensor head and analog output current | | | | PNP open-collector transistor (2 outputs) ・ Maximum source current : 50 mA ・ Applied voltage : 30 V DC or less (between comparative output and +V) ・ Residual voltage : 1 V or less (at 50 mA source current)
| NO / NC, selectable by key operation | EASY mode / Hysteresis mode / Window comparator mode | Minimum 1 digit (variable)(however, 2 digits when using psi unit) | Within ±0.2 % F.S. | 5 ms, 10 ms, 25 ms, 50 ms, 100 ms, 250 ms, 500 ms, 1,000 ms, 5,000 ms, selectable by key operation | Incorporated | ・ Output current : 1 to 5 V DC ・ Zero point : within 1 V ±0.5 % F.S. (excluding DPH-L113V) within 1.364 V ±0.5 % F.S. (DPH-L113V) ・ Span : within 4 V ±0.5 % F.S. ・ Linearity : within ±0.1 % F.S. ・ Output impedance : 1 kΩ approx.
・ Output current : 4 to 20 mA ・ Zero point : within 4 mA ±1.0 % F.S. (excluding DPH-L113V) within 5.455 mA ±1.0 % F.S. (DPH-L113V) ・ Span : within 16 mA ±1.5 % F.S. ・ Linearity : within ±0.1 % F.S. ・ Load resistance : 250 Ω (max.)
| Input voltage range : 1 to 5 V DC (over rated pressure range) | ON voltage : 5 V to +V DC OFF voltage : 0.6 V DC or less, or open Input impedance : 10 kΩ approx. Input time : 1 ms or more
| Orange LED (Comparative output 1 operation indicator, comparative output 2 operation indicator : Lights up when each comparative output is ON ) | IP40 (IEC) | -10 to +50 ℃ +14 to +122 ℉ (No dew condensation or icing allowed), Storage : -10 to +60 ℃ +14 to +140 ℉ | 35 to 85 % RH, Storage : 35 to 85 % RH | | 500 V AC for one min. between all supply terminals connected together and enclosure | 50 MΩ, or more, with 500 V DC megger between all supply terminals connected together and enclosure | 10 to 500 Hz frequency, amplitude 3 mm 0.118 in or maximum acceleration 196 m/s2, in X, Y and Z directions for two hours each (when panel mounting bracket is mounted : 10 to 150 Hz frequency, amplitude 0.75 mm 0.030 in or maximum acceleration 49 m/s2, in X, Y and Z directions for two hours each) | 100 m/s2 acceleration in X, Y and Z directions for three times each | Within ±0.5 % F.S. (ambient temperature range based on +20 ℃ +68 ℉) | | | Enclosure : PBT (glass fiber reinforced), LCD display : Acrylic, Mounting threaded part : Brass (nickel plated), Switch part : Silicone rubber | Connector | Total length up to 100 m 328.1 ft (less than 30 m 98.4 ft when conforming to CE marking) is possible with 0.3 mm2, or more, cable. | | | Net weight : 25 g approx. (excluding connector attached cable), Gross weight : 140 g approx. | Pressure unit label: 1 set | Where measurement conditions have not been specified precisely, the conditions used were an ambient temperature of +20 ℃ +68 ℉. The values specified above are applied only to the controller. | |